Signet 2759 pH/ORP System Tester

0
0

pH/ORP


Signet 2759 pH/ORP System Tester

The Signet 2759 pH/ORP Simulator is a battery-powered millivolt generator that simulates pH values of 4, 7, and 10, plus ORP values of ±700 mV.  This device is useful as a troubleshooting aid and for general verifi cation of system operation. It is not a substitute for periodic system calibration with pH buffers or test solutions.  

Accessory adapter cables (sold separately) enable the 2759 to connect directly to Signet 2760 preamplifi ers, or 2751 pH/ORP Smart Sensor Electronics.  The adapters include a selector switch for pH (3K or PT1000 Temperature Compensation) or ORP simulation. The switch triggers automatic sensor-recognition software in Signet pH/ORP instrumentation. 

Advantages:

•    Battery-powered millivolt generator
•    Simulates pH and ORP values
•    High impedance input simulates preamplified signal
•    Verifies system functionality
•    Compatible with 2751 and 2760 preamplifiers
•    Connects to all Signet instruments
•    Verifies preamplifier or instrument electronics


Features:


A)    Power OFF button
B)    Output simulation buttons and indicators Simulate pH and ORP output at fi xed values: pH 4, pH 7, pH 10,  -700 mV and +700 mV. Pressing any one of these buttons turns the 2759 on.
C)    Low battery indicator
D)    High Ω switch Adds 1000 MΩ resistance in series with output. Simulates high impedance of pH electrodes. Used to verify proper preamplifi er operation.
E)    Adapter cable Use PN 3-2759.391 for use with the 2751 or 2760.
F)    Mode selector switch Trigger automatic sensor recognition software in Signet pH/ORP instrumentation.  The three-way toggle switch positions are:
•     Top = 1K for a Signet 8900/9900/9950 instrument or 2751 Smart Sensor Electronics.  Compatible with Pt1000 or 3KΩ temperature element.
•    Middle = 10K for ORP simulation.
•     Bottom = 3K for older Signet instruments.  Requires a 3KΩ temperature compensation input.