Thermo Orion 2210AM Ammonia Analyzer

0
0

Ammonia


Thermo Orion 2210AM Ammonia Analyzer

When system efficiency is paramount, ammonia monitoring is a critical measurement. High ammonia levels can coat internal process components, leading to decreased efficiency, diminished safety, damage and costly downtime. Introducing our new Thermo Scientific™ Orion™ 2210AM Ammonia Analyzer, designed to provide continuous and online measurement of free ammonia to ensure process treatment optimization. Thermo Scientific™ analyzers are designed to provide fast, accurate and reliable online ammonia detection – optimized for any application that does not require Free ammonia in the system.


Advantages:

  • Low reagent consumption due to a highly optimized, state of the art design
  • Wide range of detection – 0.03 - 30 mg/L provides a complete picture of ammonia treatment process
  • Large, easy-to-read display allows for easy viewing in all types of lighting conditions
  • Simple menu navigation – an easy-to-read and understand menu structure
  • Compact size capable of being panel- or wall-mounted
  • Intelligent Sample Awareness – intelligent algorithm that senses the sample to keep the analyzer in a ready state, while offline, for up to 30 days

Fields of Application:

  • Drinking water
  • Surface water

 

Instrument:

  • The 2210AM analyzer measures ammonia concentrations using a salicylate colorimetric method
  • Measurement is semi continuous and is <15 minutes.
  • The range of measurement is between 0.03 and 30 mg/L as N.
  • The detection limit is 0.03 mg/L, in the range of 0.03 - 30 mg/L as N for 5-40 °C.
  • Repeatability is <±2 % or ±.0.05 mg/L, whichever is greater, in the range of 0.03 - 30 mg/L as N.
  • Analyzer is capable of auto calibration.
  • Analyzer takes blank reading before every measurement and compares and corrects the measurement.
  • Analyzer provides two isolated 4 - 20 mA outputs and four programmable alarm relays. Relays are rated for 2 A @ 250 V AC.
  • Analyzer has an hourly and daily maintenance mode that will move solutions and purge lines to keep the analyzer ready while offline for up to 30 days.
  • Analyzer has a heated sample cell with sample presence detection
  • Display is a digital graphics LCD with backlight.
  • Sample delivery requirement is between 50 and 100 mL/min.
  • Analyzer is capable of grab sample analysis.
  • Analyzer can be wall or panel mounted.
  • Analyzer fluidics cabinet is constructed to NEMA 4X/IP65. Enclosure is constructed of fiberglass with clear polycarbonate windows.
  • Power requirement is 100 – 240 V AC, 110 W, 50/60 Hz, auto-detection.
  • Analyzer has cTUVus approvals to meet UL61010-1 and CSA C22.2 No. 61010-1 certifications and FCC Class A requirement.
  • Analyzer shall meet EC Directives 2014/30/EU and 2014/35/EU.
  • Analyzer is aware of sample availably.